CCS50 Analog chlorine dioxide sensor
Field of application
CCS50 is a robust, low-maintenance sensor for chlorine dioxide measurement. It provides stable and fast measured values in water, process water and utilities. The sensor ensures efficient disinfection even at minimum water volumes and helps to achieve highest water quality and safety, to avoid overdosing of chlorine dioxide or to ensure the absence of ClO2 in beverage plants and reverse osmosis. CCS50 ensures continuity for the installed base of analog chlorine dioxide measuring points.
CCS50 is an analog chlorine dioxide sensor for skid builders and end customers. It measures reliably in:
-
Drinking water - to ensure sufficient disinfection
-
Cooling systems or towers - to avoid pathogen growth
-
Process water - for hygienic and safe packaging and bottling of products
-
Beverage plants - to secure the absence of chlorine dioxide
Benefits
-
Fast response time (t90 < 15 s) provides an accurate process view and enables prompt reaction to process changes as well as efficient process control.
-
Increased process safety: precise and long-term stable measurement ensures consistent process monitoring and allows for individually adapted disinfectant dosing.
-
Flexible installation: sensor can be installed in the CCA151 and CCA250 flow assemblies or in immersion assemblies. Measurement virtually flow independent at flow velocities above 5 l/h (CCA151), 30 l/h (CCA250) or above 15 cm/s (immersion).
-
The low-maintenance, amperometric sensor reduces the cost of ownership of the measuring point, particularly compared to colorimetric measuring systems.
-
Connection to the Liquisys chlorine transmitter ensures continuity for the installed base of analog chlorine dioxide measuring points.
-
Measuring principle
Chlorine dioxide
-
Application
Guarantees reliable disinfection in drinking water
Prevents the growth of biofilms and pathogens in cooling water
Ensures food quality
Ensures the absence or presence of chlorine dioxide in Utilities -
Characteristic
Amperometric measurement of dissolved chlorine dioxide
-
Measurement range
Trace: 0 to 5 mg/l ClO2
Standard: 0 to 20 mg/l ClO2 -
Measuring method
Closed, membrane covered measuring cell
Reduction of chlordioxide at the cathode -
Design
Closed amperometric 2-electrode measuring cell with PVDF membrane
-
Material
Sensor shaft: POM
Membrane: PVDF
Membrane cap: PVDF
Sealing ring: FKM -
Dimension
Diameter: 25 mm (0.98 inch)
Length: 126 mm (4.96 inch) -
Process temperature
0 to 55 °C (32 to 130°F), non-freezing
-
Process pressure
Max.1 bar (max. 14.5 psi)
-
Temperature sensor
10k NTC
-
Connection
Analog with fixed cable