Home    Instrumentation    Optical analysis    SS2100a gas analyzer
SS2100agasanalyzer01

SS2100a gas analyzer

Exceptionally reliable for measuring trace gas components

Field of application

The SS2100a gas analyzer measures accurate trace gas components (H2O, CO2, H2S, NH3, and C2H2) in natural gas and hydrocarbon processing applications using tunable diode laser absorption spectroscopy (TDLAS) technology. It requires little maintenance and does not need recalibration or periodic replacement parts. The SS2100a is certified for ATEX Zone 2.

  • Gas analysis: ppb level measurement of H2O & NH3, ppm level measurement of H2O, H2S, CO2, NH3, and C2H2, and percent (%) level measurement of H2S and CO2

  • Natural Gas: production, storage, transportation, distribution

  • Refinery: hydrogen recycle, catalytic reforming, flaring, cracking

  • Gas Processing: fractionation, sour/acid gas removal

  • LNG: liquefication, custody transfer

  • Petrochemical: cracking, acetylene converter, caustic wash tower

Benefits

  • Accurate, real-time measurements

  • Low cost of ownership; no consumables; virtually maintenance free and reliable in harsh environments

  • Measurements not hampered by wet-up (absorption) or dry-down (desorption)

  • Dependable with greater repeatability than surface-based sensors

  • Analog and serial outputs for remote monitoring

  • Available for the following measurements: H2O, CO2, H2S, NH3, C2H2

  • ATEX Zone 2 certified

  • Measuring principle

    TDLAS

  • Product Headline

    An advanced gas analyzer for hydrogen sulfide (H2S) measurement. The SS2100a uses tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the analyte without coming into physical contact with the stream.

  • Channels

    1

  • Analyte and Measurement ranges

    H2S (Hydrogen Sulfide):
    0 to 10
    0 to 1000 ppmv
    0 to 5000 ppmv
    0 to 5% volume

  • Measured Variables

    Concentration
    Cell Pressure
    Cell Temperature

  • Ambient Temperature range

    -20 to 50°C (-4 to 122°F)
    Optional: -10 to 60°C (14 to 140°F)

  • Operating Pressure range

    Inlet Pressure: 140 to 350 kPa (20 to 50 psig)
    Sample Cell: 800 to 1200 mbara or 950 to 1700 mbara (optional)

  • Analyzer Wetted materials

    316L stainless steel
    FKM O-Rings
    Glass

  • Power supply

    120 or 240 VAC ± 10 %, 50 to 60 Hz; 60W max (with 2 solenoids) - electronics enclosure
    120 or 240 VAC, 50 to 60 Hz - standard; 100W or 200W max for heated systems - sample cabinet

  • Communication

    Analog Output: Two 4-20mA Isolated, 1200 ohms at 24 VDC max load (measurement value)
    Analog Input: One 4-20mA Isolated, 1200 ohms at 24 VDC max load (process pressure)
    Serial: RS232C and Ethernet
    Protocol: Modbus Gould RTU or Daniel RTU or ASCII
    Digital Outputs: 5, Concentration Alarm, General Fault, Validation Fail, Validation 1 Active, Validation 2 Active
    Digital Inputs: 2, Flow Alarm, Validation Request

  • Housing materials

    Electronics: Copper-free Aluminum
    Sample System Enclosure: 304 or 316 stainless steel

  • Hazardous area approvals

    ATEX Zone 2

  • Degree of protection

    IP66

  • Product safety

    CE