Home    Instrumentation    Optical analysis    Raman Rxn5 process analyzer
RamanRxn5processanalyzer01

Raman Rxn5 process analyzer

Turn-key process Raman analyzer for quantitative chemical composition measurements

Field of application

The Raman Rxn5 process analyzer is an industry-leading, turn-key, laser-based analyzer powered by Kaiser Raman technology. It provides quantitative chemical composition measurements using Raman spectroscopy. The design of the Raman Rxn5 incorporates customer requirements for serviceability and hazard-area certification, provides a compact footprint, and minimizes utilities consumption. The high optical efficiency of the analyzer makes it ideal for typical gas-phase sample measurements.

Process measurement points:

  • Steam methane reformers

  • Gasifiers (coal, petcoke, waste, biomass)

  • Shift converters

  • Hydrotreaters

  • Hydrocrackers

  • CO2 removal

  • Finished product synthesis loops

Typical industries:

  • Hydrogen production/purity, HyCO, Hydrogen recycle,

  • Methanol production

  • Ammonia production

  • Gas turbine fuel feed

  • Synthetic natural gal/IGCC power plants

  • LNG baseload customer transfer, LNG rundown to storage tanks, LNG mixed refrigerant optimization

Benefits

  • Non-destructive analysis of gases, including homonuclear diatomics (H2, N2, O2)

  • Compact, smaller than most wall-mount GCs

  • Can replace up to four traditional analyzers with a single base unit and four probes

  • Rugged with minimal utility requirements, no sample transport, and no consumables (columns, valves, gases, pumps)

  • Minimal analyzer technician time with maximum technician safety (no toxic gases are transported to the analyzer)

  • Measures four streams simultaneously for fastest multi-channel results

  • -20°C to + 50°C operating environment

  • Measuring principle

    Raman spectroscopy

  • Spectral coverage

    150-3425 cm-1 (532 nm)

  • Spectral resolution

    5 cm-1 (532 nm)

  • Channels

    Up to four channels

  • Temperature

    Operating:
    -20 to 50 °C / -4 to 122 °F
    Recommended storage:
    -30 to 60 °C / -22 to 140 °F

  • Relative humidity

    0 to 90 %, non-condensing

  • Input voltage

    90-264 VAC, 47-63 Hz standard

  • Power consumption (W)

    <300 (max)
    <300 (typical start-up)
    <200 (typical running)

  • Warm up time (minutes)

    120

  • Unit dimensions (width x height x depth)

    457 x 834 x 254 mm
    18.00 x 32.84 x 10.00 inch

  • Weight

    61.2 kg / 135 lbs

  • Sampling probe compatibility

    Raman Rxn-30

  • Automation interface

    Modbus (TCP/IP or RS485)

  • Installation options

    IP 56 wall-mounted, sealed/purged enclosure

  • Hazardous area certifications

    ATEX, CSA, IECEx

  •