Raman Calibration Tool
Field of application
The Raman Calibration Tool (RCT) offers laboratory quality calibration in a field-capable tool, providing high accuracy critical for quantitative analysis. Coupled with wavelength (x-axis) and laser calibrations that are integrated into Raman analyzers, the Raman Calibration Tool provides the highest level of accuracy for even the most demanding Raman measurements. Used routinely, the RCT protocol facilitates calibration transfer between laboratory instruments and on-line process analyzers.
-
LNG truck loading
-
LNG rundown to storage tanks
Benefits
-
Robust calibration protocol
-
Excellent system accuracy and unparalleled instrument precision
-
Minimal technical support required
-
Incorporates NIST Standard Reference Material (SRM); Compatible with ASTM E2911
-
Facilitates easy calibration transfer
-
Chemical methods and models instrument-independent
-
Simple wizard-based protocol
-
Measuring principle
Raman spectroscopy
-
Temperature
Operating:
-20 to 50 °C / -4 to 122 °F
Recommended storage:
15 to 25 °C / 59 to 77 °F -
Relative humidity
Recommended storage: < 10 % humidity, non-condensing
-
Unit dimensions (width x height x depth)
83 x 66 mm / 3.27 x 2.60 inch
-
Weight
0.54 kg / 1.20 lbs
-
Spectral intensity reference
NIST SRM glass
-
Spectral intensity output repeatability (at time of certification)
< ±2 %
-
Long term spectral uncertainty (at any wavelength)
±2.85 %
-
Spectral intensity calibration range
534.5 to 694.0 nm